Resistance to dynamic and static loading of the implant mounts on its respective implant


1 Posgraduate Program in Prosthodontics, Dental Faculty, University of, Costa Rica
2 National Laboratory of Materials and Structural Models, University of, Costa Rica
3 Department of Restaurative Dentistry, Dental Faculty, University of, Costa Rica
4 Posgraduate Program in Prosthodontics, Dental Faculty; Department of Restaurative Dentistry, Dental Faculty, University of, Costa Rica

Correspondence Address:
Dr. Tatiana Vargas-Koudriavtsev
Posgraduate Program in Prosthodontics, Dental Faculty, University of Costa Rica, Department of Restaurative Dentistry, Dental Faculty, University of Costa Rica
Costa Rica
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jdi.jdi_4_21

Rights and Permissions

Introduction: Implant restorations should endure a variable range of forces over a long period of time. Some commercial brands offer the implant together with an accessory called “implant mount” or “implant holder,” which might be used as a temporary abutment. However, scientific literature in the use of implant holders as abutments for restorations is scarce. Objectives: The purpose of this in vitro study was to compare the load at which implant holders of Implant Direct® and Zimmer® fail under static compression after being subjected to fatigue, and to compare the gap produced between the implant–holder complexes after dynamic loading. Materials and Methods: The test protocol was based on the recommendation of ISO 14801. Five implant–implant holder assemblies of each brand were subjected to dynamic loading. A load of 250 N was applied at 5 × 106 cycles and at 15 Hz stress frequency (Eden Prairie, MN, USA). The gap (μm) at the interface was measured postfatigue using scanning electron microscopy (S-3700N, HITACHI, Japan), and afterward, static loading was applied and the maximum load (N) after the point of failure was established. Implant–definitive abutment complexes were used as controls. Data were analyzed by means of a central tendency measurement test Mann–Whitney U-test (nonparametric). Results: There was no difference between both the implant holder groups (P ≤ 0.05); however, a slight trend of greater resistance was observed for the Zimmer® group. The gap in the interface was greater for Implant Direct® implants, but the difference was not statistically significant. Conclusion: No significant differences were found in terms of the maximum load under compression or the interface gap after the dynamic loading in the two experimental groups.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed12916    
    Printed36    
    Emailed0    
    PDF Downloaded136    
    Comments [Add]    

Recommend this journal